Assay 2 User Notes

320

0.9

1.0

1.0

1.1

1.1

0

1 2

3 4

5 6

7 8

9 10

Run No.

Toluene 10 ppm

Benzene 2.5 ppm

Ethylbenzene 10
ppm

Figure 32 Peak Area Stability for Repetitive Sampling of 10 ppm Toluene,
2.5 ppm Benzene and 10 ppm Ethylbenzene

0.9

1.0

1.0

1.1

1.1

0

1

2 3

4

5

6

7 8

9 10

Run No.

o-Xylene 10 ppm

m-Xylene 10 ppm

Figure 33 Peak Area Stability for Repetitive Sampling of 10 ppm m-Xylene and
10 ppm o-Xylene

Response Drift

Detector response drift is caused by drift in the effective UV lamp intensity over
time. Although

Explorer maintains a constant visible lamp intensity, it is

possible that the ionizing spectral lines, located outside the visible range, shift in
intensity over time.