Explorer Operator Manual

125

Alternatively, you can use the auto cycling features. When you start auto cycling,

Explorer will begin a run on the next column to be analyzed. The next analysis will

be run on the next enabled column. The instrument determines column selection.
If you are using auto cycling, you must use loop injections. See Auto Cycling
Operation on page

148.

NOTE: Columns must be enabled on

Explorer.

Starting a Run

To start a run:

1. Ensure

Explorer is set up correctly.

Carrier gas must be supplied to the instrument.

Explorer must be displaying a

Ready status.

2. Select the Assay tab and ensure the correct assay is loaded.

3. Ensure

Explorer is on and ready for operation and then send the information to

Explorer by clicking

Send to

Explorer

.

4. Select the Instrument tab and select the column on which to perform a run.

Open the Analyzer menu and select the column or right-click on the column
graphic in the instrument status box to open a column selection pop-up menu.
Selecting

Column V

will start a TVOC run.

5. Connect a sample to the sample inlet port or allow

Explorer to draw a sample

of room air. If you are performing a syringe injection prepare the syringe
sample.

6. Press

Start

to begin the run.

Explorer will begin a run according to the method for the selected column.
A timer at the bottom of the instrument status box shows you the analysis time
and progress of the run.
The chromatogram will develop in the graph box. As peaks are integrated they
will be displayed in the peak report.