INFICON Explorer Portable Gas Chromatograph User Manual

Page 232

Advertising
background image

Calibration

230

3. When the analysis is complete you must store the new point.

If you are storing a number of compounds then you may want to print the
analysis for ease of reference. Use a PC and the SiteChart software to print the
analysis.

4. Press Enter Menu, then LIBRARY, LIST, EDIT, and STORE, and enter the

peak number to be stored in the box that opens, then press Enter Menu to
confirm the peak number.
Explorer will display the matching compound name from the current library.

5. Press Enter Menu to confirm.

NOTE: If the identified compound is not the compound you are trying to store, then you are

using the wrong library. Press Enter Menu, then LIBRARY and use the left/right
arrow key to select the correct library.

6. Use the up/down arrow keys to select the point you want to add and press Exit.

7. Use the arrow keys to enter the correct concentration and press Enter Menu.

8. Use the arrow keys to enter the window size.

You may want to decrease the window for higher concentrations.

9. Press Enter Menu to complete the process.

You have now stored the compound at two concentrations.

10. Repeat these steps for all compounds you want to store and for all

concentration points.

11. Connect a supply of calibration gas and press Start Stop to begin an analysis.

12. When the analysis is complete, review the compound list and ensure all your

compounds are present.