Explorer Operator Manual

191

19. Repeat this procedure for the other compounds to be stored in this library.

20. If possible run the compounds at more than one concentration. Use STORE to

store up to five concentration values for a compound.
Best response results will be obtained this way.

21. Once all the peak information has been entered, press Enter Menu, DISPLAY,

GC, GRAPH, REINT to re-integrate the chromatogram.
Explorer will use the new library information to identify the peaks.

22. If your compounds of interest require another column option, select the next

column and run another analysis. Store the compounds as outlined above.

23. Repeat this procedure for a third column if necessary.