Explorer Operator Manual

44

2. Install the plastic clip to

Explorer using one of the screws provided with the

probe.

3. Remove the middle screw and install the clip.

Keep the screws that you removed in the sample probe bag.

4. Tighten the screws down to ensure the

Explorer case is securely closed.

5. Connect the probe to

Explorer's sample in port. Use the compression fitting

supplied with the probe.

Using the Probe

To extend the probe, loosen the knurled nut at the top of the wand, pull the
extension out and then tighten the knurled nut to lock the extension in place.

To lengthen the probe further, loosen the lower knurled nut, pull the second
extension out and then tighten the knurled nut to lock the extension in place.

To reduce the probe, loosen both knurled nuts and push the extensions back
into the probe, then tighten the knurled nuts down to lock the extensions in
place.

CAUTION

When using the telescoping sample probe, NEVER aspirate liquids or
solids as they will damage the instrument.

Removing the Probe

1. Turn

Explorer off.

2. Disconnect the probe from

Explorer's sample in port.

3. Remove the mounting clips.