Explorer Operator Manual

271

Cause: Pieces of cored septum have fallen down inside the injection port.

Action: Remove the injection port and clean it as outlined in Cleaning the
Injection Port Assembly on page

249.

Problem: Peak appears but is not recognized

Cause: Compounds have drifted from original assay retention time values.

Action: Update the compound library and once the library has been updated save the
new values. You must do this for each column/detector pair.

Cause: Peak is not stored in the library.

Action: Ensure the compound peak is stored in the library.

Cause: Retention time window is too small.

Action: Adjust the window percent as discussed on page

115.

Cause: The sample was injected onto the wrong column.

Action: Ensure you have selected the correct column for your application. Refer to
the user notes provided with your assay.

Cause: Flow rates are not set correctly.

Action: Ensure the pressure settings are correct for your application. Refer to the
user notes supplied with your assay.

Problem: Peaks appear too slowly

Cause: Pressure is too low resulting in low detector flow rates.

Action: Check for a pressure fault and ensure the internal carrier gas cylinder is full,
or ensure the pressure setting is correct for the selected method.