Battery pack care, Excessive heat and cold, Adequate ventilation – INFICON Explorer Portable Gas Chromatograph User Manual

Page 20

Advertising
background image

Warnings and Safety Information

18

Battery Pack Care

The battery pack contains

nickel-metal hydride. This battery pack must be

disposed of properly. Check with local codes for special disposal instructions.

Charge the battery pack using the AC adapter provided with, or identified for use
with, this product only, in accordance with the instructions and limitations specified
in this manual.

Excessive Heat and Cold

The normal operating temperature is 0 C to 40 C (32 F to 105 F) with 0 to 95%
(non-condensing) relative humidity.

Do not expose the instrument to intense sunlight for prolonged periods. This may
result in a rise in instrument temperature and loss of oven control. Exposure to
excessive heat may also result in instrument contamination. Low temperatures will
cause decreased battery life and may also result in loss of oven control. The plastics
may become fragile and shatter at extreme low temperatures.

Adequate Ventilation

If compound threshold limit values (TLVs) are exceeded, you should use a gas bag

and the gas bag adapter for sampling and provide adequate ventilation during

calibration. When using the gas bag adapter tubing, do not kink or stress the

tubing. The gas bag and gas bag adapter are available in both

field kits for assay 1,

assays 2, 10, and ethylene, and calibration kit for assays 1, 2, 10, and ethylene

(MX380318, MX380348, MX380338, MX380358, and MX380337).