Explorer Operator Manual

38

WARNING

Do not modify or disassemble the carrier gas filling station. If you have
any problems, the unit must be returned to

INFICON for repair or

replacement.

To fill the internal cylinder:

1. Connect the filling station fitting to the gas cylinder.

You must specify a CGA 590, or CGA 580 when ordering the filling station.

Figure 3 Carrier Gas Refill Adapters (CGA 590 and CGA 580)

2. Turn the filling station valve handle so that it points away from the cylinder.

3. Open the cylinder valve and check that there are no leaks.

4. Place

Explorer on a sturdy flat surface within easy reach of the flexible filling

hose.

3-way valve

Quick connect

Relief valve

Flow restrictor

Flexible hose