INFICON Explorer Portable Gas Chromatograph User Manual

Page 227

Advertising
background image

Explorer Operator Manual

225

12. Press DISPLAY, GC, and CMPD to view the compound list.

Since you are viewing the ECD output, the compound list will show the ECD
compounds.

13. When the analysis is complete, review the compound list and ensure all of your

ECD compounds are present.

14. Press Enter Menu, then LIBRARY and CAL, and when prompted to calibrate

using the last analysis, select YES.
If

Explorer can match all the compounds in the last analysis with compounds

you have selected for calibration then your ECD library information will be
updated accordingly.

15. Switch to the PID view by pressing DISPLAY, GC, and DET.

16. Review the compound list and ensure all of your PID compounds are present.

17. Press Enter Menu, then LIBRARY and CAL, and when prompted to calibrate

using the last analysis, select YES.
If

Explorer can match all the compounds in the last analysis with compounds

you have selected for calibration then your PID library information will be
updated accordingly.

NOTE: If a calibration fault is displayed see Calibration Fault starting on page

266 for

information about correcting the fault

NOTE: If you need to calibrate a second, or third, column, select the next column and

disable the previous one. Ensure only one column and both detectors are selected. If
you are calibrating Column B, select PID Column C and ECD Column C. Calibrate
the column as above, and if you are using more than one column, enable the
columns once you have calibrated each one of them.

Advertising