Method Development
174
2. In the Method box, click
Config
.
3. Enter the desired
Pressure
.
4. Click
OK
to accept the changes and close the dialog boxes.
The change you have made applies to the currently selected column only. If
you need to change the setting for another column, select the next column and
change the pressure.
5. Save the assay.
Detector State
An ECD is available for
Explorer configured for Assay 1. You must have the
ECD installed to access the ECD options. In the Instrument tab, select the PID,
the ECD or both.
You can set the PID to high sensitivity for low concentrations, or for compounds
that do not respond well to a PID. Select low sensitivity for high concentrations.
To change the detector sensitivity:
1. Select the Assay tab.
2. In the Method box, click
Config
.
3. Enter the desired
PID State
and/or
ECD State
.
4. Click
OK
to accept the changes and close the dialog boxes.
The change you have made applies to the currently selected column only. If
you need to change the setting for another column, select the next column and
change the detector state.
5. Save the assay.