Calibration procedure, Single detector – INFICON Explorer Portable Gas Chromatograph User Manual

Page 223

Advertising
background image

Explorer Operator Manual

221

Calibration Procedure

Single Detector

1. Select a column.

2. Press Enter Menu, then SETUP, CONFIG, and COLUMN.

3. Use the arrow keys to move through the list and press Enter Menu to select

PID Column A, B or C or ECD Column A, B or C.
Ensure only one column/detector combination is selected.
There are two libraries for each column. The PID library or the ECD library is
automatically selected when you choose a column/detector combination.

4. Press Exit to close the box.

5. Press Enter Menu, then LIBRARY, and LIST.

6. Press the left/right arrow to select the library that matches your selected

column.
The status line at the top of the screen will tell you which library you are using.
Compounds with a check mark are currently selected for calibration.

7. Use the up/down arrows to move through the list of compounds, and then to

enable or disable a compound, move to the compound name and press EDIT
and CALCMPD.
This will toggle the compound on and off. If the compound is not selected
remove it from the library. If the compound is not selected, its response and
retention time will be updated ratio-metrically.