Multiple point calibration curve, Line segment calibration curve – INFICON Explorer Portable Gas Chromatograph User Manual

Page 228

Advertising
background image

Calibration

226

Multiple Point Calibration Curve

NOTE: Changes made to the library using

Explorer directly, cannot be saved to disk. The

changes will be lost when a new method is loaded.

Multiple point calibration may be used to improve the accuracy of

Explorer’s response over a wide concentration range.

Within a library, each compound may be stored at up to 5 different concentration
points. The first point is always (0, 0), and you can store 4 additional concentration
points. You can also select one of four calibration curves; line segment, linear fit,
quadratic fit or cubic fit.

When you perform a calibration, the calibration curve is updated based on the single
concentration point contained in the calibration gas.

Line Segment Calibration Curve

If one concentration point is stored,

Explorer draws a line from this (0, 0) to the

single point and uses the line as the calibration curve. If the compound is stored at
a second, higher concentration, sample concentrations are calculated using the first
sensitivity when the concentration is below the lower concentration. For
concentrations above the lower value,

Explorer uses a sensitivity value calculated

from the slope of the line connecting the two concentration response points.