INFICON Explorer Portable Gas Chromatograph User Manual

Page 264

Advertising
background image

Maintenance and Troubleshooting

262

3. Ensure

Explorer is on and ready for operation.

4. Connect a sample to the Sample In port and begin a run.

5. When the analysis is complete, select the

Assay

tab and use

Open

from the

File menu to open the file that was just run.
The chromatogram will be displayed in the window

6. Double-click on the peak in the chromatogram and enter the concentration.

7. Normally, you will use a linear calibration curve, select

Linear

and SiteChart

will update the calibration curve to incorporate the new point.

8. If you have more than one compound at a different concentration then repeat

the procedure for the other peaks in the chromatogram.

9. Perform another run at another concentration and store the compound and

repeat this procedure for all available concentrations.

10. If you need to remove a concentration point, select the point in the

area/concentration table and click

Erase

.

During calibration, the single point that is contained in the calibration gas is used to
update the curve.