INFICON Explorer Portable Gas Chromatograph User Manual
Page 264
Maintenance and Troubleshooting
262
3. Ensure
Explorer is on and ready for operation.
4. Connect a sample to the Sample In port and begin a run.
5. When the analysis is complete, select the
Assay
tab and use
Open
from the
File menu to open the file that was just run.
The chromatogram will be displayed in the window
6. Double-click on the peak in the chromatogram and enter the concentration.
7. Normally, you will use a linear calibration curve, select
Linear
and SiteChart
will update the calibration curve to incorporate the new point.
8. If you have more than one compound at a different concentration then repeat
the procedure for the other peaks in the chromatogram.
9. Perform another run at another concentration and store the compound and
repeat this procedure for all available concentrations.
10. If you need to remove a concentration point, select the point in the
area/concentration table and click
Erase
.
During calibration, the single point that is contained in the calibration gas is used to
update the curve.