Explorer Operator Manual

309

Assay 2 Standard Method

The Assay 2 standard method, Assay2.app, has been customized for and is
installed in your

Explorer. The typical method parameters for Assay 2 are listed

below as are the library compounds for each column and typical retention times
and detectable concentration ranges for this method. These values will vary on

your instrument.

NOTE: For maxi

mum accuracy at low concentrations, we recommend that you verify

Explorer’s integration, and reintegrate manually, if necessary.

Concentrations outside these specified ranges, especially at the lower end, can be
obtained by adjusting the injection time, PID state and integration parameters.

Settings for Assay 2

Column B

Method

Carrier Gas

Air

Detector

PID

Analysis Time (S)

1300

Oven Temperature ( C)

60

Pressure (psi)

12

PID State

Low Sensitivity

Injection Method

Inject Method

Loop Injection

Inject Time (S)

5

Pump Time (S)

20

Backflush Time (S)

300