INFICON Explorer Portable Gas Chromatograph User Manual

Page 16

Advertising
background image

Warnings and Safety Information

14

WARNING

To reduce the risk of fire or injury to persons, read and follow these
instructions:

All calibration, maintenance and servicing of this device, including
battery charging, must be performed in a safe area away from hazardous
locations.

For replacement battery pack use only

PN 951-402-G1 .

Do not dispose of the battery pack in a fire. The cell may explode. The
battery pack contains 20% cadmium (Cd) by weight. This battery pack
must be disposed of properly. Check with local codes for special disposal
instructions.

Do not open or mutilate the battery pack. It may be toxic if swallowed.

Exercise care in handling battery packs in order not to short the terminals
with conducting materials such as rings, bracelets and keys. The battery
or conductor may overheat and cause burns.

Do not defeat proper polarity orientation between the battery pack and
battery charger.

Charge the battery pack using the charger provided with or identified for
use with this product only in accordance with the instructions and
limitations specified in this manual. For battery charger use only Part No.
MX380302 (110 Volts AC).

Do not alter this cylinder or the high pressure components of this device, in any
way.

Do not over pressurize this cylinder (device).

Do not expose pressurized cylinder (device) to temperatures in excess of 54 °C
(130 °F).

Cylinders exposed to a fire or heated to temperatures in excess of 177 °C
(350 °F) should be condemned or hydrostatically tested prior to filling.

Do not remove the warning label.

Advertising