Explorer Operator Manual

269

Problem: Sensitivity is too low

Cause: Calibration gas pressure is low.

Action: Check the calibration gas pressure and replace the tank if necessary.

Cause: Incorrect calibration gas.

Action: Ensure the calibration gas contains all of the compounds listed in the library.
Each library compound must be identified as a calibration compound.

Cause: Calibration gas is not pressurized properly.

Action: Ensure the delivery pressure is maintained at no more than 14 psig.

Cause: Calibration gas has deteriorated.

Action: Replace the calibration gas tank.

NOTE: Some calibration mixes are not stable for extended periods. Check with your

specialty gas supplier for details.

Cause: UV lamp is failing.

Action: Replace the UV lamp, as detailed in Replacing the UV Lamp or Installing

a New One on page

245.

Cause: The ECD is contaminated.

Action: Set the detector temperature to 80 C and increase the pressure to 14 psi.
Allow the detector to purge under these conditions for at least 8 hours as
detailed in Purging the ECD on page

254, or contact INFICON.