To set the injection volume:
1. Select the Assay tab.
2. In the Method box, click
Config
.
3. Click
Inject
in the Configure dialog box.
4. Ensure
Syringe
is selected.
5. Enter the
Inject
volume.
6. Click
OK
to accept the changes and close the dialog boxes.
The change you have made applies to the currently selected column only. If
you need to change the setting for another column, select the next column and
change the injection volume.
7. Save the assay.
NOTE:
Explorer is not designed to accept liquid injections, so water samples are analyzed
using the headspace above standards. You may obtain commercially prepared
standards or you can prepare the headspace standards.
CAUTION
Do not inject liquid samples into
Explorer or you will damage it.