Explorer Operator Manual

171

To set the injection volume:

1. Select the Assay tab.

2. In the Method box, click

Config

.

3. Click

Inject

in the Configure dialog box.

4. Ensure

Syringe

is selected.

5. Enter the

Inject

volume.

6. Click

OK

to accept the changes and close the dialog boxes.

The change you have made applies to the currently selected column only. If
you need to change the setting for another column, select the next column and
change the injection volume.

7. Save the assay.

NOTE:

Explorer is not designed to accept liquid injections, so water samples are analyzed

using the headspace above standards. You may obtain commercially prepared
standards or you can prepare the headspace standards.

CAUTION

Do not inject liquid samples into

Explorer or you will damage it.