Removing library compounds – INFICON Explorer Portable Gas Chromatograph User Manual
Page 236
Calibration
234
Removing Library Compounds
Compounds that are not required as part of your application should be removed from
the library. This may speed analysis time and will reduce the possibility of
misidentification due to co-elution. Refer to the user notes supplied with your assay
for details.
NOTE: You must make modifications to the assay in the SiteChart software. If you create
your custom assay on the instrument you cannot save it.
1. Open a copy of the assay.
Do not modify your original assay.
2. To remove compounds from the library, press Enter Menu, then LIBRARY,
LIST, EDIT, and STORE.
3. When prompted for a Peak No., enter 0 to erase the compound from the library.
4. Make all the changes required and then save it as your custom application.