149

Auto cycling is the only way to cycle through multiple columns as required for an
analysis if you are operating

Explorer from SiteChart. If you click

Start

on the

Instrument tab or select the

Start

and

Stop

commands from the Analyzer menu,

SiteChart will begin an analysis on the selected column. If you click

Start

again,

the analysis will again be on the selected column. To switch columns, you must
select another column and then click

Start

.

To configure your instrument for auto cycling operation:

1. Ensure

Explorer is set up correctly. Carrier gas must be supplied to the

instrument. It must be displaying a ready status.

2. Select the Assay tab and ensure the correct assay is loaded.

3. Select the Instrument tab and select the column on which to begin auto cycling.

4. Open the Analyzer menu and select the column or right-click on the column

graphic in the instrument status box to open a column selection pop-up menu.

5. Connect a sample to the sample inlet port or allow

Explorer to draw a sample

of room air.

6. Open the Analyzer menu and select

Setup

.

7. Enter a delay time.

This is the time between runs. A minimum delay of 1 minute is required to
enable

Explorer to stabilize the baseline.

8. Enter the number of runs required.

If your method requires all three columns and you need 2 complete analyses,
enter 6. After 6 runs, 2 on each column,

Explorer will turn auto analyzing off.

If you want the instrument to sample continuously enter 0. If you are analyzing
continuously, ensure that you have connected

Explorer to the AC adapter and

an external supply of carrier gas. Check the

Auto Cycle

box and then click

OK

.

Explorer will begin auto cycling.