Explorer Operator Manual

261

12. If the compound peak is the only peak within the retention time window, click

Update

or double-click on the peak to open the Update dialog. If the

information in the box is not correct, enter the correct data and click

OK

.

OR

If the compound peak is not within the retention time window or if there are
other peaks in the retention time window, drag the center, blue, dashed line to
the desired peak and the retention time will be updated automatically.

13. Repeat this procedure for each library compound.

14. Select

Re-integrate

from the Analyzer menu.

Each compound should be correctly identified.

15. If you are using another column for your application, repeat this procedure for

the next column.

16. Save the assay and then send it to

Explorer.

17. Using SiteChart, switch to the

Instrument

tab and run an analysis on each

column.

18. Ensure all compounds are correctly identified and then use these analyses to

calibrate

Explorer.

Updating the Calibration Curve

After you have updated the compound retention times you can also update the
calibration curve for each library compound. It is not necessary to update the
calibration curves, but multiple point calibration will provide the most accurate
response.

1. In the SiteChart software, select the

Assay

tab.

2. Select the column/detector combination required for your compound.

Ensure your compound is already listed in the library.