Explorer Operator Manual

50

Immediate Availability

If you use

Explorer every day, you will most likely want the instrument ready to use

immediately. If you do not use the instrument regularly, but still require immediate
availability then you must store the instrument as described below.

Leave

Explorer turned on.

Calibrate once every 24 hours.
Calibration should be performed before beginning operation of

Explorer,

as noted in Calibration on page

210.

Connect the instrument to a carrier gas cylinder, or refill the carrier gas
cylinder every 6 - 8 hours.

Connect the instrument to the AC adapter when not is use.

Availability Within 30 Minutes

If you use

Explorer less frequently, you may want to use the following procedure for

storing the instrument. This procedure allows you to store the instrument up to 5
days. When you are ready to use

Explorer, allow approximately 30 minutes for the

instrument to warm up. Better analysis result may be obtained if you allow the
instrument to warm up for 1-2 hours.

Turn

Explorer off.

Connect the instrument to an external cylinder of carrier gas.

Connect the instrument to the AC adapter.

When you are ready to use

Explorer, turn it on and allow 30 minutes for

the instrument to warm up and stabilize. When all faults have cleared,
calibrate

Explorer for your application.