Explorer Operator Manual

239

NOTE: Each filter is protected by a piece of blue plastic. Remove the plastic before

installing the filter in

Explorer.

4. Remove the Teflon/Polypropylene filter and install the new filter

(MX396015).

5. Place the filter so that the Teflon side is facing down in the filter housing and

the mesh side is facing

Explorer.

Handle the filter disk only by the edges. The mesh may be damaged or
contaminated by excessive handling. Use forceps or tweezers if possible.

6. Replace the filter housing.

NOTE: Calibrate all GC and TVOC modes before returning to normal operation.

CAUTION

NEVER operate

Explorer without an inlet filter.