2. When the run is complete, review the chromatogram and peak report to ensure

that all compounds have been detected and that the reported concentrations are
correct.

3. Select the Assay tab and ensure all library concentrations are correct.

4. Open the Analyzer menu and select

Calibrate

.

SiteChart will calibrate

Explorer. The Explorer library will be updated

accordingly.

Explorer is now calibrated and ready for operation.

The information in the SiteChart library will not be updated. Each time you
load a new assay and download it to the instrument, you must calibrate the
instrument in order to update the current library information.

Service

The

Service

command displays a window that shows the

Explorer display. This

can be used to troubleshoot instrument problems remotely. This command allows
you to access all

Explorer functions from your computer.

NOTE:

Explorer cannot be powered up from this window. If Explorer is powered down, then

you must turn it on by pressing the On/Off button on the instrument.

The buttons control

Explorer in the same manner as the keys on Explorer.

Communication Menu

The Communications menu allows you to setup communications between the
computer and the instrument.

The communication pull-down menu has the following options:

Setup

allows you to configure which serial port will be used to connect the

instrument to the remote computer.

Dial Out

allows you to dial out to the specified number and establish

communication with a remote computer. This feature will allow you to dial out
to a remote instrument from SiteChart.