Maintenance and Troubleshooting
276
Problem: Peak cannot be integrated
Cause: Min Area or Min Height values are set too high.
Action: Lower the values of Min Area or Min Height. If you are not using auto-
integrate, turn the option on and re-integrate the chromatogram.
Problem: A value of zero is displayed, even though a valid number was
entered
Cause: Exit was pressed, instead of Enter Menu to confirm the entry.
Action: Type the number into the input box, ensure it is correct and then press Enter
Menu.
Cause: Exit was pressed before the value was complete.
Action: Type the number into the input box, ensure it is correct and then press Enter
Menu.
Cause: The value you have entered is beyond the allowable range for the instrument
Action: Type a valid number into the input box, ensure it is correct and then press
Enter Menu.