INFICON Explorer Portable Gas Chromatograph User Manual

Page 278

Advertising
background image

Maintenance and Troubleshooting

276

Problem: Peak cannot be integrated

Cause: Min Area or Min Height values are set too high.

Action: Lower the values of Min Area or Min Height. If you are not using auto-
integrate, turn the option on and re-integrate the chromatogram.

Problem: A value of zero is displayed, even though a valid number was
entered

Cause: Exit was pressed, instead of Enter Menu to confirm the entry.

Action: Type the number into the input box, ensure it is correct and then press Enter
Menu.

Cause: Exit was pressed before the value was complete.

Action: Type the number into the input box, ensure it is correct and then press Enter
Menu.

Cause: The value you have entered is beyond the allowable range for the instrument

Action: Type a valid number into the input box, ensure it is correct and then press
Enter Menu.