Maintenance of the injection port, Replacing the injection port septum – INFICON Explorer Portable Gas Chromatograph User Manual

Page 246

Advertising
background image

Maintenance and Troubleshooting

244

Maintenance of the Injection Port

Replacing the Injection Port Septum

A worn out septum causes leakage, which will result in poor reproducibility and an
unstable baseline. When an injection is made, the baseline will suddenly drop below
zero as the needle is pushed against the septum. This will also occur if the septum is
improperly seated in the injection port.

A silicone septum with a Teflon face (MX380320) is recommended for use
with

Explorer. This type of septum will last for about 10 injections depending

on the syringe needle.

Septum Retainer

Injection Port

Cover

Septum

Figure 15 Replacing Septum

1. Move the instrument to an area where the ambient air contains no VOCs.

2. Unscrew the black septum retainer.

3. Use a syringe needle to withdraw the old septum.

Advertising