Method Development

204

3. Press Exit to close the box. Press Enter Menu to close the menu.

4. Press Enter Menu, SETUP, CONFIG, METHOD, METH, and AUTO.

5. Use the arrow keys to select on and press Enter Menu. Press Exit to close the

box.

6. Press Enter Menu, DISPLAY, GC, GRAPH and REINT.

The signal will be re-integrated using the new integration settings.

7. Press DISPLAY, GC, PEAK to see if the peaks have been integrated

accurately.

8. If you need to change the settings for the other two columns, repeat this

procedure for each column.

9. When you have changed the settings for each column, enable all the columns

required for your method.

Integrator Delay

The integrator delay is used to delay the start of integration after the analysis begins.
It is used to prevent integration of peaks which may appear as a result of pressure
changes during sample injection, especially when large syringe injections are made.
If you have more than 25 peaks in an analysis, integrator delay may be used to
determine which 25 peaks will be identified. If you are interested in the peaks in the
middle or at the end of the chromatogram, set the integrator delay so that the first
peaks are not integrated.

1. Press Enter Menu, SETUP, CONFIG, COLUMN.

2. Select the column for which you want to alter the integration settings. Disable

the other columns.

3. Press Exit to close the box. Press Enter Menu to close the menu.

4. Press Enter Menu, SETUP, CONFIG, METHOD, METH, and DELAY.