Audio options, Alarm audio, Injection audio – INFICON Explorer Portable Gas Chromatograph User Manual

Page 69

Advertising
background image

Using

Explorer

Audio Options

Alarm Audio

If any of the identified peak concentrations exceed the alarm level, stored in the
library, then the audible alarm may sound at the end of the analysis.

You have three options for the audio alarm. You can turn it off and there will be no
audible indicator if an alarm occurs. Turn the audio off if you are auto-analyzing
or if you are controlling

Explorer from a remote location through SiteChart. You

can turn the audio alarm on, so an audible alarm will sound until the alarm is
acknowledged. You can also set the alarm to sound momentarily and then shut off.

To change the audio option:

1. Press Enter Menu, then SETUP, CONFIG, and AUDIO.

2. Use the arrow keys to select the desired option from the Set Audio box and

press Enter Menu to confirm your selection.

3. Press Enter Menu to close the menu.

Injection Audio

You can configure

Explorer to sound an audible indicator when the sample is

injected onto the column. This injection beep is most useful for syringe
injections.

1. Press Enter Menu, then SETUP, MODE, and INJECT.

2. Press the speaker icon.

3. Use the arrow keys to select off and press Enter Menu and then Exit.

67