Maintenance schedule – INFICON Explorer Portable Gas Chromatograph User Manual
Page 238
Maintenance and Troubleshooting
236
Maintenance Schedule
WARNING
Do not perform any maintenance procedures on
Explorer in
hazardous locations.
Explorer requires regular maintenance to keep it operating efficiently. Replacement
of some items will vary with usage, the following table shows the suggested
maintenance periods:
Item/Operation
Frequency
Replace injection port septum
Every 10 injections
Replace inlet filter
Once a week
Clean UV lamp window
Once a week or as required
Update assay
As required
Wipe Test the ECD (Contact your local
authority for regulations.)
Every 6 months (USA),
every year (Canada)
Replace UV lamp
Once a year or as required
Replace calibration gas (commercial cylinder)
Once a year or as required
Purge internal carrier gas cylinder
As required
Condition column
As required
Charge battery pack
As required