Maintenance and Troubleshooting

236

Maintenance Schedule

WARNING

Do not perform any maintenance procedures on

Explorer in

hazardous locations.

Explorer requires regular maintenance to keep it operating efficiently. Replacement

of some items will vary with usage, the following table shows the suggested
maintenance periods:

Item/Operation

Frequency

Replace injection port septum

Every 10 injections

Replace inlet filter

Once a week

Clean UV lamp window

Once a week or as required

Update assay

As required

Wipe Test the ECD (Contact your local
authority for regulations.)

Every 6 months (USA),
every year (Canada)

Replace UV lamp

Once a year or as required

Replace calibration gas (commercial cylinder)

Once a year or as required

Purge internal carrier gas cylinder

As required

Condition column

As required

Charge battery pack

As required