INFICON Explorer Portable Gas Chromatograph User Manual

Page 118

Advertising
background image

SiteChart

Software

116

Once you have stored multiple points you can select the best curve to fit the points.

If one concentration point is stored, SiteChart draws a line from this (0, 0) to the
single point and uses the line as the calibration curve. If the compound is stored at a
second, higher concentration, sample concentrations are calculated using the first
sensitivity when the concentration is below the lower concentration. For
concentrations above the lower value, SiteChart uses a sensitivity value calculated
from the slope of the line connecting the two concentration response points.

Figure 6 Linear Calibration Curve
If a third, higher concentration is stored, then the concentrations above the second
value are measured using a sensitivity value calculated from the slope of the line
connecting the second and third concentration response points. You can store up to
five concentration points.

A quadratic calibration curve fits a quadratic function to the points you have stored.
In order to use the quadratic calibration curve, you must store at least 2 or more
points. The curve will always start at (0, 0) and go through at least 2 points you have
stored and (0,0).

In order to obtain the most accurate results you should store a point that is at the top
end of the desired concentration range. If you are interested in the range 10 to 100
ppm, store a point at 100 ppm. Data past this point will be extrapolated based on a
tangent through the last stored point.

Advertising