Method Development

192

Analysis Settings

NOTE: Changes made to the method, using

Explorer directly cannot be saved to disk. The

changes will be lost when a new method is loaded.

Analysis Time

Analysis time is the duration of the analysis. You must set the analysis time so that
all compounds of interest clear the column and pass into the detector. Details of how
to set the analysis time are provided in the user notes specific to your assay.

To change the analysis time:

1. Press Enter Menu, SETUP, CONFIG, COLUMN.

2. Select the column for which you want to alter the analysis time. Disable the

other columns.

3. Press Exit to close the box and then Enter Menu.

4. Press Enter Menu, SETUP, CONFIG, METHOD, TIMING, and ANLZ.

5. Use the arrow keys to enter the new analysis time.

6. Press Enter Menu.

7. If you need to change the analysis time for the other two columns, repeat this

procedure for each column.

8. When you have changed the analysis time for each column, enable all the

columns required for your method.

Advertising