Explorer Operator Manual

231

13. Press Enter Menu, then LIBRARY and CAL, and when prompted to calibrate

using the last analysis, select YES.
If

Explorer can match all the compounds in the last analysis with compounds

you have selected for calibration then your Library information will be updated
accordingly and the instrument will be ready for operation.

NOTE: If a calibration fault is displayed see Calibration Fault starting on page

266 for

information about correcting the fault.