INFICON Guardian EIES Controller User Manual

Page 107

Advertising
background image

4 - 15

IP

N 07

4-

51

7-

P1

D

Guardian Co-Deposition Controller Operating Manual

On the General tab, select Internet Protocol (TCP/IP) and press the Properties
button. See

Figure 4-14

.

Figure 4-14 Internet Protocol (TCP/IP) Properties

Select the button for Use the following IP address, enter the IP address and
Subnet mask shown in

Figure 4-15

, and press OK. With this selection, the PC is

assigned an IP address to use when communicating with the Guardian.

The Guardian is shipped from INFICON with a pre-assigned address of
192.168.1.200. To communicate directly with the Guardian from a PC, the PC must
also be assigned an address that starts with 192.168.1 but cannot be set to
192.168.1.200. The example in

Figure 4-15

uses the address 192.168.1.100 for

the PC.

Advertising