3 qcm tab – INFICON Guardian EIES Controller User Manual

Page 83

Advertising
background image

3 - 31

IP

N 07

4-

51

7-

P1

D

Guardian Co-Deposition Controller Operating Manual

Gas Cal: INFICON's gas compensating sensors provide a measurement of the
emission caused by background gases. The gas calibration factor is used to null
this portion of the signal from the main flux sensor measurements. If the EIES rate
reading with no evaporation is negative, decrease the Gas Cal. parameter until the
rate averages near zero. If the rate reading is positive with no evaporation,
increase the Gas Cal. parameter until the reading averages near zero.

Enable Gas Sensors: Click to select the gas compensating feature. Requires gas
compensating (GC) sensors.

3.8.3 QCM Tab

NOTE: See the SQM-242 Card or Q-pod Manual for detailed QCM installation and

troubleshooting instructions.

None: No QCM inputs are used.
SQM_242: INFICON SQM-242 card is used for QCM measurements. DLL shows
the status of the SQM-242 card's Windows drivers. DLL return values of 9XX
indicate a card installation error. The firmware revision of the installed SQM-242
and SAM-242 cards are also listed. A value of 0 indicates the card is not installed.

Figure 3-24 Edit System QCM Tab