Guardian Co-Deposition Controller Operating Manual

Current: A percent of each xChan Detector current is subtracted, before any
material calibration, tooling factors, etc. have been calculated. The advantage
of this method is that a change in the xChan calibration factors does not impact
the amount of xChan compensation.

Rate: A percent of each xChan Detector rate is subtracted, after material
calibration, tooling factors, etc. have been calculated. This method allows
recalibration of an xChan to alter the amount of xChan compensation.

3.6.4 Ratio Display Grid

In addition to cross channel calibration, the xChan tab of the Edit Layer dialog box
allows setup of the Main Screen Ratio Display.

Figure 3-16 Ratio Display

Channel Multiplier: The channel reading is multiplied by this value before the ratio
is calculated. Useful for scaling or calibrating to a desired value. 0.00 disables the
function.

Ratio Channel #: Channel used for (denominator of) ratio calculation.

Ratio Channel Multiplier: Multiplies ratio channel reading by this amount.
0.00 disables the function.