Guardian Co-Deposition Controller Operating Manual

7.3.4.7 G

Description: Get PMT readings.

Readings are returned as a comma delimited string of average A to D counts (Cts)
and number of A to D conversions (Cnv) use to calculate the average Cts. The
readings for each PMT module are separated by a semi colon. Readings are only
returned for modules with a new reading available. The format of the returned
string when a new reading is available for PMT modules 1, 2, and 8 is:

OnCtsPMT1,OnCnvPMT1,OffCtsPMT1,OffCnvPMT1,GasCtsPMT1,GasCnvPMT1;

OnCtsPMT2,OnCnvPMT2,OffCtsPMT2,OffCnvPMT2,GasCtsPMT2,GasCnvPMT2;

;;;;;OnCtsPMT8,OnCnvPMT8,OffCtsPMT8,OffCnvPMT8,GasCtsPMT8,GasCnvPMT8

Use the Slope and Intercept values calculated using the C command to calculate
the current for a PMT module:

OnCurrentPMT# = 2000nA - (Slope(PMT#) * OnCtsPMT#) + Intercept(PMT#)

OffCurrentPMT# = 2000nA - (SlopePMT# * OffCtsPMT#) + InterceptPMT#

GasCurrentPMT# = 2000nA - (SlopePMT# * GasCtsPMT#) + InterceptPMT#

Finally, the measured PMT current is calculated by:

CurrentPMT# = (OnCurrentPMT# - OffCurrentPMT#) -

(GasCurrentPMT# - OffCurrentPMT#)

Parameters: None

Example: G

Returns: G;992445,99,992492,99,992462,100;
992445,99,992492,99,992462,100;;;;; 992445,9

7.3.4.8 H

Description: Sets PMT bias voltage. Values from 0 V to 1250 V are possible.

See Note in

section 7.3.3.2 on page 7-9

re "safe mode" timeout.

Parameters: PMT#,Voltage

Example: H1, 750 (set PMT1 bias voltage to 750 V)

Returns: H