Time Setpoint Relay: This relay will close when Run Time equals the Time
Setpoint.

Channel Relays 1 to 8: These relays control EIES sensor shutters. They close
during shutter delay and deposition.

Source Relays 1 to 8: These relays control the Shutter that covers the deposition
source. They close during deposition.

QCM Relays 1 to 4: These relays control QCM sensor shutters. These relays
close during shutter delay and deposition.

3.8.4.2 Input Assign:

To assign a deposition event to an input, click the Input #, then select the desired
event from the drop-down menu. Click the Read button to read the current state of
the highlighted input. A brief description of each input event follows:

Start Process: Triggering this input is the same as pushing the Start Process
button.

Stop Process: Triggering this input will abort the process.

Start Film: Triggering this input starts the current film.

Stop Film: Triggering this input will stop the current film.

Emission 1/2 Interlock: Set this input to inhibit turning on emission from the front
panel.

Emission 1/2 On: Turns sensor 1 or 2 emission On

Emission 1/2 Off: Turns sensor 1 or 2 emission Off

End Channel 1 to 8: Stops deposition phase and goes to idle phase

.

Calibrate Channel 1 to 8: Starts calibration of the selected channel.

NOTE: Input 5 is hardware assigned to the General Emission interlock function.