6.3.3 Tooling Factor
Tooling Factor adjusts for the difference in material deposited on the quartz sensor
versus the substrate. Tooling may be less than or greater than 100% as shown in
Figure 6-5 Tooling Value
1
Place your substrate and a new quartz sensor in their normal position.
2
Set Tooling to an approximate value; Density as determined above;
Z-Ratio =1.
3
Deposit approximately 1000 to 5000 Å of material.
4
Use a profilometer or interferometer to measure the substrate's film thickness.
5
The correct Tooling Factor is calculated by:
[11]
[12]
where:
U
q
= 32 Gpa
Substrate
Substrate
Tooling
Over 100%
Tooling
Under 100%
Tooling
Actual
Tooling
Approx
Thickness
Actual
Thickness
QCM
------------------------------------------
=
Z
D
q
* U
q
D
m
* U
m
1 2
=