3 tooling factor, Tooling, Thickness – INFICON Guardian EIES Controller User Manual

Page 131

Advertising
background image

6 - 7

IP

N 07

4-

51

7-

P1

D

Guardian Co-Deposition Controller Operating Manual

6.3.3 Tooling Factor

Tooling Factor adjusts for the difference in material deposited on the quartz sensor
versus the substrate. Tooling may be less than or greater than 100% as shown in

Figure 6-5

.

Figure 6-5 Tooling Value

1

Place your substrate and a new quartz sensor in their normal position.

2

Set Tooling to an approximate value; Density as determined above;
Z-Ratio =1.

3

Deposit approximately 1000 to 5000 Å of material.

4

Use a profilometer or interferometer to measure the substrate's film thickness.

5

The correct Tooling Factor is calculated by:

[11]

[12]

where:

U

q

= 32 Gpa

Substrate

Substrate

Tooling

Over 100%

Tooling

Under 100%

Tooling

Actual

Tooling

Approx

Thickness

Actual

Thickness

QCM

------------------------------------------

=

Z

D

q

* U

q

 D

m

* U

m

1 2

=