10 m, 11 n, 12 s – INFICON Guardian EIES Controller User Manual

Page 153

Advertising
background image

7 - 15

IP

N 07

4-

51

7-

P1

D

Guardian Co-Deposition Controller Operating Manual

7.3.4.9 K

Description: Sets the output voltage of control outputs 1 to 6 to a value between
-10 V to +10 V. The command is in DAC counts with 0 = -10V, 32767= 0 V, and
65535=+10 V

See Note in

section 7.3.3.2 on page 7-9

re "safe mode" timeout.

See the W command below for outputs 7 & 8.

Parameters: Out#,Counts

Example: K1,32767 (set Output 1 to 0 V)

Returns: K

7.3.4.10 M

Description: Reads the state of the digital inputs. The response is a comma
delimited string of 13 ones or zeros. The first 12 values correspond to each of the
12 inputs. The 13th value is unused. A value of one is input true (+V) , 0 is input
false (0V).

Parameters: None

Example: M

Returns: M1,0,1,1,1,1,1,1,1,1,1,1,1 (Input 2 is false, all others are true)

7.3.4.11 N

Description: Sets the state of the twelve relay outputs. The command is a comma
delimited string of 12 ones or zeros. One is relay closed, 0 is replay open.

Parameters: Relay1,Relay2,Relay3,Relay4,Relay5,..........,Relay12

Example: N1,0,1,1,1,1,1,1,1,1,1,1 (Relay2 is open, all others are closed)

Returns: N

7.3.4.12 S

Description: Turns sensor emission on/off.

See Note in

section 7.3.3.2 on page 7-9

re "safe mode" timeout.

1: Sensor 1 standard sensor
2: Sensor 1 gas sensor
3. Sensor 2 standard sensor
4. Sensor 2 gas sensor

Parameters: Sensor#, On/Off

Example: S2,1 (turns on Sensor 1 gas sensor)

Returns: S