Rockwell Automation 2755 Enhanced Decoder Series B User Manual

Page 93

Advertising
background image

7–5

Configuration: Scanner Control, Primary Match Table, Discrete I/O

Publication 2755-833

Inter-Scan Timer

Field Type Options

Edit 10 through 9999 or 0

Disables decoding for a set time interval after a valid read. The
default is 0, which disables the timer. The Inter-Scan Timer is valid
in both continuous and triggered modes of operations.

You can use the Inter-Scan Timer to:

provide a delay between decodes to enable multiple labels, that
are not necessarily unique, to be read within the same package

enable multiple decodes of the same label for verification
purposes

The Inter-Scan Timer accepts a value between 10 and 9999.

Important: To ensure that package labels are read before the
No-Read Timer expires, set the Inter-Scan Timer to a value that is
less than the No-Read Timer setting.

A valid read starts the timer. See Figure 5.2. The decoder decodes a
scan within Field #1, which starts the Inter-Scan Timer and
immediately stops decoding. The decoder waits for the timer to
expire (or timeout) to resume decoding.

Use the following equations to create a delay between decodes:

W (cm or inches) x 1000 (msec/sec)

Line Speed (cm/sec or inches/sec)

Approximate Inter-Scan Timer Value (msec)

=

Equation 1:

Equation 2:

Maximum Inter-Scan Timer Value

S (cm or inches) x 1000 (msec/sec)

Line Speed (cm/sec or inches/sec)

Approximate Inter-Scan Timer Value (msec)

=

Minimum Inter-Scan Timer Value

You may have to experiment with several settings since factors such
as label quality determine when the first read occurs.

ATTENTION: Set output durations in multiples of 5 milliseconds. Actual output will have a worst-case precision

of +0 to –5 milliseconds, varying with the timing of the output. Durations that are not set in multiples of 5
milliseconds will be rounded up to the nearest multiple of 5 and function within the same +0 to –5 millisecond
precision.

For example, a setting of 25 milliseconds will result in a duration of from 20 to 25 milliseconds. A setting of 23
milliseconds will result in the same duration range.

Advertising