©BÜCH Labortechnik AG

6 - Starting up and shutting down the instrument

108

Room temperature should be constant.

Analyzer and peripherals must be protected from drafts and direct heat.

An air-conditioned room is advantageous as the analyzer emits heat.

Take suitable action to avoid electrostatic discharge. Electrical discharges can lead to
malfunctions of the instrument.

Specific requirements

The following lists specific requirements to be met by the installation site

Type

Specific requirements

Room

Mode of protection IP 20.

Dry room.

Well ventilated.

Instrument setup

Only on heat-resistant surface.

A stable lab table is suitable as well.

Space required including PC and printer

1.30 m x 0.55 m (W x D).

Space required by instrument alone

0.48 m x 0.55 m x 0.55 m (W x D x H).

Clearance from instrument to wall

At least 20 cm clearance all round so that the fans will
work.

Ambient temperature

Minimum: + 15°C.

Maximum: + 35°C.

Mains voltage

100–230 Volt AC ± 10 %; 50–60 Hz.

Connection 1,4 kVA.

Mains voltage and frequency must match the
specifications on the instrument's type plate.

Power connections

2 power sockets and socket strip with 4 sockets to
connect peripherals or 5 sockets.

Protection class I, protective conductor connection.

Fusing 16 A delay action.

Gas supply

Gas supply must be available nearby the instrument.

Connection to device by means of a pressure reducer.

Pressure reducer CO

2

and O

2

, two-stage each (2nd

stage 0-3 bar), if possible with stop valve.

Necessary gas connection downstream of the
pressure reducer: 2 x clamp ring screw connection for
6 mm tube diameter.

The complete gas intake line must use gas-loss-free
components (copper, stainless steel, etc.).

Waste gas lines

The waste gas lines "WASTE" and "REF" must
discharge into the open or into a vent, since otherwise
the emitted carrier gas CO

2

will be enriched in the

laboratory.

The end of the waste gas line must discharge into the
open at a location protected from the wind.

Gases and chemicals to be provided

Gases and chemicals to be provided

The following lists the gases and chemicals required.