O_3757 – BUCHI DuMaster D-480 User Manual

Page 235

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

235

Drying tube filling (large)

The following picture shows the fillings of the drying tube:

Legend:

1 Empty space
2 Cotton (10 mm)
3 Sicapent® (314 mm)

Safety instruction

Strictly observe the safety instruction "Filling the drying tube" under Warning notes during operation
(on page 24).

Filling the drying tubes

Proceed as follows:

 Disconnect both tubes.

 Remove the glass tube.

 Detach the screw-on caps.

 Leave approx. 5mm space (1) at the end of the tube.

 Plug approx. 10 mm of cotton (2) into the tube so that it comes to rest above the empty space

(3).

 Fill in enough Sicapent® (3) to leave approx. 15mm space at the top of the tube. To do so, tap

the tube wall with a piece of wood or cork so as to prevent cavities. If the Sicapent® cannot be
tapped properly, also a glass rod for slight compacting can be used.

 Clean any Sicapent® off the end of the tube.

 Plug approx. 10 mm of cotton over the Sicapent®.

 Replace the screw-on caps and tighten them on the glass tube by hand.

 Push the glass tube between the clamps and close the tubes.