Importing weighing data, Prioritizing urgent samples, O_1107 – BUCHI DuMaster D-480 User Manual

Page 134: O_583

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

7 - Using the instrument

134

... "Auto-decrement the number in the sample name": Works the same way as auto-
increment, only the figure in the name is decremented.

... "Increment/decrement interval": Defines the amount the number in the sample name is
increased/decreased.

 If you have activated one of the options "Auto-increment the number in the sample name" or

"Auto-decrement the number in the sample name" you have to set the interval value in the
textbox "Increase/decrease interval" by means of the arrow buttons.

 Check the checkbox next to the appropriate options.

Effect of the option...

... "Fill method names to stop marker": The method name input for the first sample is
entered for all subsequent samples up to the stop marker.

... "Fill daily factors to last sample": The daily factor input for the first sample is entered for
all subsequent samples.

... "Fill blank values to last sample": The blank value input for the first sample is entered for
all subsequent samples.

 Click OK. This closes the dialog and saves the settings.

Importing weighing data

Importing weighing data

Proceed as follows:

 Open the "Open" dialog by selecting File > Export/Import > Import.

 Select the "Balance Files (*.dat)" file type from the "File type" dropdown listbox.

 From the dropdown listbox select the directory containing the weighing data file.

 Select a file and click Open.

 Decide:

If you import into...

... an empty sample memory: The weighing data are saved to sample memory starting at
No. 1. Save the document. The import is finished.

... a sample memory that is not empty: The "Reading Balance file" dialog is displayed. Go to
the next step.

 Select the desired radiobutton in the dialog:

If you...

... select "Append samples as from the current weighed sample": The weighing data are
stored starting after the current sample weight.

... select "Create new document": The weighing data are saved in a new document.

 Click OK. The weighing data are imported and displayed in the sample field.

 Save the document. The import is finished.

Prioritizing urgent samples

Prioritizing urgent samples

Proceed as follows:

 Open the "Swap samples" dialog in the operating software by selecting Edit > Swap.

 Enter the numbers of the samples to swap in the textboxes in the dialog. Then click OK.