Defining the balance / pc interface, O_1445 – BUCHI DuMaster D-480 User Manual

Page 87

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

5 - Work performed by the system administrator

87

Defining the balance / PC interface

Safety instruction

Strictly observe the safety instruction "Improper changes" under Warning notes during operation
(on page 24).

Defining the balance / PC interface

Proceed as follows:

 Open the "Device configuration" dialog by selecting Options > Configuration.

 Select the "DEVICE.INI" tab. Then select the "Balance Port" section by clicking the arrow next to

the "Section" dropdown listbox.

 In the listbox "Items" select the entry that you want to change and click Edit. The "Configure key

value" dialog is displayed:

 Enter the desired value in the "Value" field and click OK. The "Comment modification" dialog is

displayed.

 Enter a justification in the "Comment modification" dialog and click Close.

 If necessary, repeat the steps mentioned above to modify further settings.

 Click OK in the "Device configuration" dialog. Your settings are saved and the dialog is closed. A

message window appears which informs that the modification will only become operative after
the instrument software has been left and re-started.

 Click OK and quit the operating software.

 Restart the operating software as required.

Dialog details

Details on the "Device configuration" dialog can be found in Device configuration (on page 320).