Lims settings (dialog), O_412 – BUCHI DuMaster D-480 User Manual

Page 311

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

10 - Menu and dialog descriptions

311

Open

Open the dialog via Options > Settings > Calculation.

Elements

The following describes the elements of the dialog.

In section "Factor determination":

Label

Element type

Meaning

sequentially

Radiobutton

The daily factor is computed sequentially.

total

Radiobutton

The daily factor is computed as a total.

manual input

Radiobutton

The daily factor can be entered manually, there will be no
calculation of the daily factor.

autom. factor
calculation

Checkbox

calculation of the daily factor starts automatically after the
end of every analysis.

separate upper/lower
calibration range

Checkbox

Determines whether valid factors shall be computed for

both calibration ranges or separate ones for each range.

In section "Blank value determination":

Label

Element type

Meaning

sequentially

Radiobutton

The blank value is computed sequentially.

total

Radiobutton

The blank value is computed as a total.

manual input

Radiobutton

The blank value can be entered manually, there will be no
calculation of the blank value.

autom. blank value
calculation

Checkbox

calculation of the blank value starts automatically after the
end of every analysis.

Remaining elements:

Label

Element type

Meaning

OK

Button

Saves new settings

Cancel

Button

Discards settings not yet saved.

Further information

Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:

Determining the daily factor (on page 145).

Determining blank values (on page 144).

Specifying the computation method for blank value and daily factor (on page 123).

LIMS settings (Dialog)

Function

This dialog is used to make various adjustments to a laboratory information and management system
(LIMS). The dialog has the following tabs:

On the "LIMS data" tab you select which data to send to the LIMS and whether they should be
sent automatically after every analysis.

On the "Communication" tab you specify whether and how to transfer data.

Advertising