Optimizing sleep and wake-up behavior, O_584, Ge 135) – BUCHI DuMaster D-480 User Manual

Page 135

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

7 - Using the instrument

135

 This closes the dialog and the samples are swapped in the sample memory.

 Manually swap samples in the carousel.

Optimizing sleep and wake-up behavior

What can you optimize?

You can set sleep/wake-up behavior so that conditioning samples in a series are measured before
the first user starts work on the instrument in the morning.

Note

Remember to prepare the appropriate conditioning samples and to sort them into the carousel after
the samples of the last measurement of the day.

Optimizing sleep and wake-up behavior

Proceed as follows:

 Open the "Sleep/wake-up functions" dialog in the operating software by selecting Options >

Settings > Sleep/Wake-up.

 The last measurement of the day must be stopped before the appended conditioning samples.

To do so, select the radio button next to "Sleep at sample No." in the "Sleep" section. In the box
next to it, enter the sample number at which you want the instrument to sleep; it corresponds to
the first conditioning sample.

 In the "Wake-up" section, enter the time and days on which to wake up the instrument.

 To do so, check the checkbox next to "Continue after wake-up if there are samples".

 Click OK. This closes the dialog and saves the settings. The conditioning samples will be

processed the next time the instrument wakes up.

Advertising