Changing the position of the carousel, Export analysis data for support, O_3568 – BUCHI DuMaster D-480 User Manual

Page 251: O_751

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

9 - Repairing the instrument

251

 Weigh the sample If the weight is not transmitted online, enter the weight by hand in the sample

view on the PC.

 Insert the new sample on the carousel.

 Open the "Set current weigt" dialog again by selecting Edit > Current weight.

 Use the arrows to set the number following the number of the last sample.

 Click OK. This closes the dialog and the cursor moves to the "Weight" column of the selected

sample number.

 If 21 CFR Part 11 functionality is enabled, the "Comment change" dialog is displayed. Enter a

justification in the "Comment modification" dialog and click Close.

Changing the position of the carousel

When is this necessary?

It is necessary to change the position of the carousel in the following cases:

If the analyzer was switched off, the carousel is no longer in zero position. Before you can start
an analysis, you must move the carousel back to zero position.

If you lose samples, it may be necessary to move the carousel to a different position.

Changing the position

Please observe the following instruction:

Caution

Inserting multiple samples into the ball valve (only for

solid operation)

If there are samples in the carousel while you are changing the position of the
carousel via System > Carousel position, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.

Before changing the position of the carousel, always remove all samples from the
carousel.

Proceed as follows:

 Remove all samples from the carousel.

 Open the "Set carousel position" dialog via System > Sample Position.

 Once you have removed all samples from the carousel, check the checkbox next to "All samples

removed from carousel".

 Decide:

If you want to...

... move the carousel to a position other than zero position, select the radio button next to
"Position" and use the arrows to set the position in the box next to it.

... move the carousel to the zero position, select the radiobutton next to "Reference run".

 Click OK. This closes the dialog and moves the carousel to the desired position.

Export analysis data for support

Export for support

The service department can draw conclusions about the condition of the analyzer from the analysis
data, particularly the graphical information. Therefore, analysis data must be exported in mdb format
for support, as this also contains graphical data.