BUCHI DuMaster D-480 User Manual

Page 32

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

3 - Product description

32

Interfaces

USB, RS232 /V24, Safety low voltage as per DIN IEC 380/VDE
0806/08.81

Dimensions and weights

The following table contains dimensions and weight of the instrument:

Reference value

Technical specifications

Dimensions

480 mm x 550 mm x 550 mm (Width x Depth x Height)

Weight

65 kg

Electrical and gas supply

The following table contains the technical specifications of the electrical and gas supply:

Reference value

Technical specifications

Mains voltage

100–230 Volt AC ± 10 %; 50–60 Hz

Power requirement

0.6 kWh/h

Connected load

1.4 kVA

Connected electrical loads and
frequency of the complete
analysis unit

Connected electrical load:
Voltage:


Frequency:

16 A
100V AC ± 10 %
120V AC ± 10 %
230V AC ± 10 %
50–60 Hz

Supply gases

Carbon dioxide:
Oxygen:

Purity: 99.995 %
Purity: 99.995 %

Consumption of supply gases

Per analysis approx. 4 litres carbon dioxide.
Per analysis approx. 0.1 to 1.0 litres oxygen.

Ambient conditions

The following table contains the technical specifications of the ambient conditions:

Reference value

Technical specifications

Protection class

Protection class I, protective conductor connection

Mode of protection

IP20, installation in dry rooms only

Noise level

55 dB (A), as per DIN 45635 section 1

Permissible ambient temperature

Maximum: + 35 °C, minimum: + 15 °C