7 - Using the instrument

139

When is routine measuring work necessary?

You must perform routine measuring work before measuring real samples:

if you have not performed any routine measuring work today

if measuring breaks exceed a duration of 2 hours.

Measuring real samples with preceding routine measuring work

Please observe the following instruction:

Caution

Inserting multiple samples into the ball valve (for solids mode only)

If there are samples in the carousel while you are changing the position of the
carousel via System > Sample Position, multiple samples may fall into the ball
valve at the same time. In this case, a service technician is required.

Before changing the position of the carousel, always remove all samples from the
carousel.

Proceed as follows:

 Move the empty carousel to zero position. To do so, open the "Set carousel position" dialog in

the operating software by selecting System > Sample Position.

 Select the radio button next to "Reference run" and click OK. The carousel moves to zero

position.

 Prepare the samples.

If necessary, prepare samples for determining the blank value.See Types of blank value
determination and their settings
(on page 143).

Prepare conditioning samples for conditioning. Settings can be found in Performing
routine measuring work
(on page 140).

Prepare the samples for determining the daily factor. Settings can be found in Performing
routine measuring work
(on page 140).

Prepare the real samples. Cf. Sample preparation instructions (on page 126).

 Insert the samples on the carousel.

 Enter the weight in [mg] in the sample view (text view) on the PC if it has not already been

transmitted online by the balance.

 Enter the sample name (max. 32 characters) in the sample view (text view):

Blank values must have the keyword defined via Options > Settings > Keywords (factory
set to "Blnk").

Daily factor or calibration samples must have the name defined in Options > Settings >
Standards
.

 Set method in the sample view (text view):

For blank samples, cf. Types of blank value determination and their settings (on page
143).

For conditioning samples, cf. Performing routine measuriement work (on page 140).

For daily factor samples, cf. Performing routine measuriement work (on page 140).

For real samples, cd. Values for oxygen dosing (on page 137).

 Go through the measuring start checklist. Cf. Checklist for blank value / conditioning and

daily factor measurements (on page 142).

 Start the analysis process in the operating software:

Select System > Auto run to start a multiple-sample analysis.

Select System > Single run to start a single-sample analysis.

 Once measurement is finished, evaluate the measurement.