6 - Heating up the furnace / checking parameters

118

 In case the analyzer has to be transported, mount the corresponding transport protection(s)

again.

Note

Note the following:

The following section is part of initial start-up. All steps described in initial start-up before "Heating up
the furnaces / checking parameters" must have already been performed. See Start-up (on page 109).

Some instrument parameters can not only be defined in the "Device parameters" dialog but also via a
method. The following general rule applies:

If not otherwise defined by a method, the settings in the "Instrument parameters" dialog apply.

If a parameter is defined differently in the method and in the "Device parameters" dialog, the
method takes priority if it is used for the measurement.

Heating up the furnace / checking parameters

Safety instruction

Strictly observe the safety instruction "Overheating of tube fillings" under Warning notes during
operation
(on page 24).

Proceed as follows:

 Open the "Device parameters" dialog in the operating software by selecting Options > Settings

> Parameters.

 Check furnace working temperatures in the "Temperatures" selection and change them if

necessary. See Setting instrument parameters (on page 50).

 Check the other parameters in the "Times" selection and change them if necessary.

 Click OK to confirm your settings. The settings are applied.

Heating up the furnace / checking

parameters