Removing the reaction tubes from the furnace, O_3747 – BUCHI DuMaster D-480 User Manual

Page 227

Advertising
background image

Operation Manual D-480

©BÜCH Labortechnik AG

8 - Maintaining the instrument

227

Removing the reaction tubes from the furnace

Process

The section is divided into the following phases:

Preparing the instrument, gas pressure

Removing reaction tubes

Safety instruction

Strictly observe the safety instruction "Removing reaction tubes" under Warning notes during
operation
(on page 24).

Requirements

Before starting work, the following requirements must be met:

Keep a tube rack ready to hold hot components.

Phase 1, Preparing the instrument, gas pressure

In order to be able to prepare the instrument for maintenance works follow the instructions

"Preparing the instrument: Gas pressure" in Preparing and following up maintenance work (on
page 171).

Furnace area

The following picture shows the details of the furnace area:

Legend:

1 Handle
2 Bayonet catch
3 Reduction tube connection
4 Post combustion tube connection

Phase 2: Removing reaction tubes

Proceed as follows:

 Open the front door.

 Unscrew the combustion tube connector. To do so, turn the bayonet catch handle (1) to the right.

 Remove the ground-in clamp on the reduction tube (3).